Clinical Evaluation of an Infrared Refracting Instrument

James W. Bizzell, J Hendricks, M. F. Goldberg, M. K. Patel, G. F. Robbins

Archives of Ophthalmology · 1974 · 10 citations · 4 references

Concepts

Abstract

Refractive errors of 567 eyes (286 patients), obtained by independent refractionists using retinoscopic and subjective techniques, were compared with those obtained by a trained technician operating the Ophthalmetron. The Ophthalmetron was incapable of producing an interpretable tracing in 26% of eyes, primarily because of poor fixation by the patients. The spherical power agreed within ± 0.5 diopter by the two objective methods in 52.6% of eyes. The cylindrical power agreed within ± 0.5 D by the two objective methods in 71.5% of eyes. The cylindrical axis measured by the two objective methods was within five degrees in 40% of eyes. The principal limitations of the infrared refracting instrument are related to (1) the lack of cooperation in very young and very old individuals; (2) disease states limiting the transparency of the ocular media, such as cataracts; and (3) aphakia.