Archives of Ophthalmology · 1974 · 10 citations · 4 references
Health ProfessionsOphthalmologyOptometryClinical OptometryGeriatric OptometryVisual OpticsAge-related Refractive ChangeRefractive ErrorRefractive Error AssessmentRetinal ExaminationVisual ImpairmentAdult Refractive Error AssessmentAge-specific RefractionClinical Reliability Of RefractionClinical RetinographyOphthalmic RefractionRefractive Error DetectionSubjective Refraction Validation
Refractive errors of 567 eyes (286 patients), obtained by independent refractionists using retinoscopic and subjective techniques, were compared with those obtained by a trained technician operating the Ophthalmetron. The Ophthalmetron was incapable of producing an interpretable tracing in 26% of eyes, primarily because of poor fixation by the patients. The spherical power agreed within ± 0.5 diopter by the two objective methods in 52.6% of eyes. The cylindrical power agreed within ± 0.5 D by the two objective methods in 71.5% of eyes. The cylindrical axis measured by the two objective methods was within five degrees in 40% of eyes. The principal limitations of the infrared refracting instrument are related to (1) the lack of cooperation in very young and very old individuals; (2) disease states limiting the transparency of the ocular media, such as cataracts; and (3) aphakia.
4
Aran Safir · Archives of Ophthalmology · 1970 · 44 citations
THE OPHTHALMETRON, PRINCIPLES AND OPERATION
Henry A. Knoll, Richard Mohrman · Optometry and Vision Science · 1972 · 31 citations
Lyon Hyams · Archives of Ophthalmology · 1971 · 23 citations
Automatic objective refraction. Report of a clinical trial.
A Safir, Henry A. Knoll, R Mohrman · PubMed · 1971 · 13 citations