Publication | Closed Access
Effects of Extracorporeal Shock Wave Lithotripsy on Cardiac Pacemakers and its Safety in Patients with Implanted Cardiac Pacemakers
53
Citations
8
References
1988
Year
(A) It is generally safe for patients implanted with standard single chamber devices in a ventricular application to undergo ESWL without modifying the pacing/sensing parameters. (B) Patients implanted with dual chamber devices who pace in the atrium should be reprogrammed to the VVI mode during ESWL. (C) Patients with piezoelectric activity sensing rate responsive single chamber pacemakers should have this feature programmed off during ESWL and, if implanted in the abdomen, probably should not undergo ESWL.