Journal of the Association for Vascular Access · 2004 · 25 citations · 20 references
MedicineSurgeryInterventional RadiologySurgical InstrumentationVascular SurgeryCatheterization CareVenous AccessCatheter Malposition PreventionCatheter SizingCatheter Tip LocalizationCentral Venous Access DevicesCentral Venous Catheter Position AssessmentCentral Venous Catheter Tip LocalizationNeedle Insertion Approach
Purpose: To validate the formula-based central venous catheter (CVC) length measurement “tailored” to individual's height and access site for predicting optimum SVC tip position.Method: A prospective study of 3 percutaneous insertion sites (PICC, SCC and JC). Formula-based “LUM'S CVC MEASUREMENT GUIDE” was used to determine the catheter length.Results: Overall, 97% (373) of the total 382 insertions were successfully placed with CVC tip in the distal SVC (SVC between carina and atrio-cava junction) location.Conclusion: The “tailored fit” formula to individual patient height is a reliable tool to predict CVC length. Appropriate catheter length can greatly reduce the guesswork and possibility of complications related to tip malposition.
20
Roger Warwick, Peter Williams, Mary Dyson et al. · British Journal of Radiology · 1985 · 4.5K citations
Accurate placement of central venous catheters
William T. McGee, Barbara L. Ackerman, LAWRENCE R. ROUBEN et al. · Critical Care Medicine · 1993 · 223 citations
Zaki Aslamy, Christian L. Dewald, John E. Heffner · CHEST Journal · 1998 · 156 citations