2012 · 10 citations · 13 references
EngineeringBioengineeringBiomedical EngineeringBiomemsNeural EngineeringProstheticsBioelectronic MedicineImplant TechnologyImplantable Biomedical DevicesImplantable Neural DevicesMedical Device SafetyNeural ProstheticsBiomedical Wireless Power TransferBlood FlowBody TemperatureElectromagnetic Field SimulationElectromagnetic Field TheoryImplantable Wireless Power TransferTransmitter DesignWireless Power Transfer For Implants
This paper derives the health risks of an inductively powered brain implanted visual prosthesis with a CAD based head model. A high efficiency self-oscillating class E transmitter circuit operating at 5 MHz has been designed to wirelessly power the implant. The electromagnetic field inside the head model has been computed using Finite Integration Technique method. Simulations illustrate that a maximum localized SAR level averaged over 10 g attains 1.91 W/kg when the power transmission is equal to 1 W, which is just below the safe level specified at 2 W/kg. Using the same power transmitter, the temperature in the brain with and without considering bio-heat factors, such as blood flow and basal metabolic rate, increases respectively with 0.918 °C and 0.293 °C above the applied body temperature of 37 °C.
13