Medical Devices Evidence and Research · 2014 · 21 citations · 33 references
MedicineCardiologyHemodynamic ManagementPhysiatryPhysical ExaminationBlood Pressure ControlBlood Pressure ManagementCardiovascular AssessmentCardiovascular MonitoringHemodynamicsVital SignsArterial Blood Pressure MeasurementBlood PressureBlood Pressure MeasurementBlood Pressure MonitoringBlood Pressure MonitorsNoninvasive Blood Pressure MeasurementSystolic Blood PressureValidation Study Design
BACKGROUND: The objective of this study was to validate the novel integration of oscillometric (Vasotens(®)) technology into a BPLab(®) ambulatory blood pressure (BP) monitoring system to measure central BP, the aortic augmentation index, and pulse wave velocity (PWV) compared with the recommended and widely accepted tonometric method. METHODS: The ARTERY Society guidelines for comparison of PWV measurement techniques were used as the basis for recruitment of 99 individuals (mean age 44±19 years, 52 males). The standard for comparison was the conventional "classic" SphygmoCor device. RESULTS: Accordance of the two methods was satisfactory (r=0.98, mean difference of 2.9±3.5 mmHg for central systolic BP; r=0.98, mean difference of -1.1±2.3 mmHg for central diastolic BP; r=0.83, mean difference of -2.6%±13% for aortic augmentation index; r=0.85, mean difference of 0.69±1.4 for PWV). CONCLUSION: The performance of Vasotens algorithms using an oscillometric ambulatory BP monitoring system is feasible for accurate diagnosis, risk assessment, and evaluation of the effects of antihypertensive drugs.