British Journal of Radiology · 1994 · 30 citations · 3 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingNoiseNoise ExposureNoise MeasurementAcoustic NoiseHigh-field Magnetic Resonance ImagingHuman Safety RisksImage Noise AssessmentMagnetic ResonanceMagnetic Resonance Imaging AcquisitionMagnetic Resonance PhysicsSafety In Magnetic Resonance Imaging
Although much has been published concerning the various safety risks associated with magnetic resonance (MR) imaging, little mention is made of the acoustic noise generated during scanning. The existing data are now out of date with regard to currently used imaging protocols and technology. Out of concern for patient safety, measurements were made of the acoustic noise generated during various scan protocols on a high field (1.5 T) Siemens MR imager. These data were compared with published data. The data show that for certain protocols, the exposure to acoustic noise falls outside safety guidelines unless ear protection is used.
3
Potential hearing loss resulting from MR imaging.
Robert E. Brummett, J. Michael Talbot, P Charuhas · Radiology · 1988 · 170 citations
Acoustic analysis of gradient-coil noise in MR imaging.
Robert Hurwitz, Sam R. Lane, R.A.I. Bell et al. · Radiology · 1989 · 98 citations
Reduction of sound levels with antinoise in MR imaging.
Arnold M. Goldman, W E Gossman, P C Friedlander · Radiology · 1989 · 76 citations