Calibration and Application of an X-ray Image Intensifier/ Charge-Coupled Device Detector for Monochromatic Macromolecular Crystallography

A. P. Hammersley, Kate Robson Brown, W.P. Burmeister, L. Claustre, Seán McSweeney, E.P. Mitchell, Jean‐Pierre Moy, Sigfrid Svensson, Andrew Thompson

Journal of Synchrotron Radiation · 1997 · 59 citations · 0 references

DOIFull text

Open access

Concepts

Abstract

Charge-coupled device (CCD)-based X-ray detectors allow data to be collected much more quickly (approximately 10 times) than with current on-line imaging-plate systems. At the ESRF, X-ray image intensifier/CCD detector systems have been developed. These have great potential as fast read-out detectors for macromolecular and other forms of crystallography. They are relatively large sensitive X-ray detectors but have two inherent weaknesses: convex detection surfaces leading to spatial distortion and non-uniformity of intensity response, and susceptibility to small changes in magnetic fields. A large improvement has been made to the accuracy obtained by non-uniformity of response calibration and correction, using fluorescence from doped lithium borate glasses. Monochromatic macromolecular crystallography demonstration experiments with external user groups have shown that high-quality results may be obtained under real experimental conditions.