Rapid Communications in Mass Spectrometry · 2006 · 18 citations · 22 references
PharmacyChemistryAnalytical ChemistryElectroanalytical MethodsMass Spectrometric AnalysisDetector SystemsElectrochemical DetectionMass SpectrometryDetectionExcitationFourier Transform Ion Cyclotron Resonance Mass SpectrometryFourier Transform Mass SpectrometrySensitivity
Separation of the functions of ion excitation and detection between different cell compartments allows for implementation of excitation and detection techniques unattainable in a single compartment of the conventional ion cyclotron resonance (ICR) cell. In particular, multi-electrode detection at a multiple of the main cyclotron frequency can be utilized without the loss of sensitivity and other negative effects. The new O-trap designed exclusively for ion detection adds an additional, internal coaxial cylinder around which ions with excited cyclotron orbits rotate. Comparison of simulated performance characteristics of the new O-trap with those of the same-size conventional cylindrical cell shows that the O-trap can provide higher sensitivity and ion capacity. Multiplexing of the O-traps can further increase the analysis speed. Future efforts will be aimed at building and testing experimentally the coaxial O-trap, including optimization of the method of ion transfer between the compartments of the cell.
22
Currents to Conductors Induced by a Moving Point Charge
W. Shockley · Journal of Applied Physics · 1938 · 1.3K citations
Michael W. Senko, Christopher L. Hendrickson, Mark R. Emmett et al. · Journal of the American Society for Mass Spectrometry · 1997 · 461 citations