Comparisons between Neural Response Imaging thresholds, electrically evoked auditory reflex thresholds and most comfortable loudness levels in CII Bionic Ear users with HiResolution™ sound processing strategies

De-min Han, Xueqing Chen, Xiaotian Zhao, Ying Kong, Yongxin Li, Sha Liu, Bo Liu, Lingyan Mo

Acta Oto-Laryngologica · 2005 · 35 citations · 4 references

Abstract

For 8 patients (53 stimulating electrodes) the correlation between tNRI threshold and M level was r=0.675 (p=0.000) and that between first NRI thresholds and M level was r=0.741 (p=0.000). On average the M-level value was 20 CU (Current Unit) lower than the first NRI threshold value and 12 CU higher than the tNRI threshold value. The M-level patterns across the electrode array overall were similar to the tNRI or first NRI threshold patterns. For 7 patients (112 stimulating electrodes) the correlation between EART and M levels was r=0.710 (p=0.000). On average the EART value was 14 CU higher than the M-level value.