Nucleic Acid Assay System for Tier II Laboratories and Moderately Complex Clinics to Detect HIV in Low‐Resource Settings

Wen Tang, Wing Huen A. Chow, Ying Li, Huimin Kong, Yi‐Wei Tang, Bertrand Lemieux

The Journal of Infectious Diseases · 2010 · 57 citations · 19 references

Concepts

Abstract

There is a clear need for an instrument-free molecular diagnostic system for detecting human immunodeficiency virus type 1 (HIV-1) RNA or DNA that can be used in developing countries. Such a test could be used for early diagnosis of HIV-1 infection during infancy and could serve as a surrogate end point for vaccine trials. We developed the IsoAmp HIV-1 assay (BioHelix Corporation), which targets the HIV-1 gag gene with use of isothermal reverse-transcription helicase-dependent amplification chemistry. The IsoAmp HIV assay uses a disposable amplicon containment device with an embedded vertical-flow DNA detection strip to detect the presence of HIV-1 amplicons. The vertical-flow DNA detection strip has a control line to validate the performance of the device and a test line to detect the analyte. The analyte is detected by a sandwich immunoassay for reporter moieties on a capture probe and a detection probe. The control line consists of the detection probe reporter moiety conjugated to the vertical-flow DNA detection strip. The preliminary limit of detection of the IsoAmp HIV assay was evaluated by testing serial dilutions of HIV-1 armored RNA (Assuragen). We found that 21 (75%) of 28 assays yielded positive results when 50 copies of HIV-1 armored RNA were input into the IsoAmp HIV reaction.