Journal of Anesthesia · 2011 · 26 citations · 15 references
Veterinary MedicineAnesthesiologySurgeryAnesthesiaAnesthetic TechniquesGeneral AnesthesiaAirway AssessmentAirway ManagementAirway SurgerySore Throat ManagementAirway Changes After SurgeryAirway Management After AnesthesiaAirway Pressure Management
PURPOSE: The cuff volume of the Cobra perilaryngeal airway (CobraPLA) is larger than that of other alternative airway devices and makes it difficult to predict the effect of cuff pressure on the perilaryngeal mucosa. We tested the hypothesis that adjustment of the cuff pressure of the CobraPLA could reduce the incidence of postoperative sore throat (POST). METHODS: After induction of general anesthesia and insertion of the CobraPLA by standardized method, the cuff pressure was set to 60 cmH(2)O (group C, n = 87) or adjusted to minimal seal-up pressure +5 cmH(2)O (group A, n = 87). The frequency and severity (0, none; 1, mild; 2, moderate; 3, severe) of throat soreness, pain, discomfort, and adverse effects were evaluated 1 and 24 h after removal of the CobraPLA. RESULTS: Incidence of moderate POST in group C was higher than that in group A (11% vs. 2%, P = 0.021) whereas the overall POST incidence was not different between the two groups (31% vs. 20%, P = 0.092). The inflated air volume of group A was different from that of group C (41 vs. 50 ml, P = 0.009). CONCLUSIONS: Adjustment of cuff pressure reduces the incidence of moderate POST after use of the CobraPLA.
15
J. Brimacombe, Lindsey Holyoake, C. Keller et al. · Anesthesiology · 2000 · 155 citations · Full text
Pharyngeal Mucosal Pressure and Perfusion
J. Brimacombe, C. Keller, F. Pühringer · Anesthesiology · 1999 · 88 citations · Full text
Medicine, Anesthesiology, Surgery +16