Analytical Chemistry · 1978 · 19 citations · 15 references
Natural SciencesChemistryEnvironmental SciencesAnalytical ChemistryChromatographic AnalysisEnvironmental ChemistryChromatographyContaminant AnalysisEnvironmental AnalysisInstrumental AnalysisLiquid ChromatographyWater QualityAnalytical Detection Of Organic PollutantsAquatic Pollutant Chemistry
A rapid method for quantitative routine determinations of traces of tetrachloroethylene in natural waters is described. Water samples of 10-ml volume are introduced directly into a reversed-phase high pressure liquid chromatography (HPLC) system (octadecylsilica, methanol/water). Detection is performed by UV absorption at 208 nm. A precision of ±2.5% and a detection limit of 0.06 μmol of tetrachloroethylene per L of water were achieved. A wide concentration range (0.06-110 μmol L<sup>-1</sup>) can be covered. The method was successfully applied in an actual water pollution case. Other possible applications of the same technique are discussed.
15