Concepedia

Publication | Closed Access

Implantation of deep brain stimulators into subthalmic nucleus: technical approach and magnetic imaging—verified electrode locations

415

Citations

50

References

2002

Year

Abstract

The authors' approach to implantation of DBS leads into the STN was associated with consistent lead placement in the dorsolateral STN, a low rate of morbidity, efficient use of operating room time, and robust improvement in motor function. The mean coordinates of the middle of the electrode array, measured on postoperative MR images, were 11.6 mm lateral, 2.9 mm posterior, and 4.7 mm inferior to the midcommissural point, and 6.5 mm lateral and 3.5 mm anterior to the center of the red nucleus. Voltage thresholds for several types of stimulation-induced adverse effects were predictive of lead location. Technical nuances of the surgery are described in detail.

References

YearCitations

Page 1