Physiological Measurement · 1998 · 129 citations · 13 references
Electrical impedance tomography (EIT) is a technique that makes it possible to measure ventilation and pulmonary perfusion in a volume that approximates to a 2D plane. The possibility of using EIT for measuring the left-right division of ventilation and perfusion was compared with that of radionuclide imaging. Following routine ventilation (81mKr) and perfusion scanning (99mTc-MAA), EIT measurements were performed at the third and the sixth intercostal level in 14 patients with lung cancer. A correlation (r = 0.98, p < 0.005) between the left-right division for the ventilation measured with EIT and that with 81mKr was found. For the left-right division of pulmonary perfusion a correlation of 0.95 (p < 0.005) was found between the two methods. The reliability coefficient (RC) was calculated for estimating the left-right division with EIT. The RC for the ventilation measurements was 94% and 96% for the perfusion measurements. The correlation analysis for reproducibility of the EIT measurements was 0.95 (p < 0.001) for the ventilation and 0.93 (p < 0.001) for the perfusion measurements. In conclusion, EIT can be regarded as a promising technique to estimate the left-right division of pulmonary perfusion and ventilation.
13
Jorge A. Wernly, Tom R. DeMeester, Peter T. Kirchner et al. · Journal of Thoracic and Cardiovascular Surgery · 1980 · 220 citations
Surgical Oncology, Medical Imaging, Advanced Lung Disease +12
Changes in the thoracic impedance distribution under different ventilatory conditions
G Hahn, I. Šipinková, F. Baisch et al. · Physiological Measurement · 1995 · 154 citations