IEEE Transactions on Neural Systems and Rehabilitation Engineering · 2012 · 13 citations · 11 references
EngineeringBiomedical EngineeringElectrical EngineeringElectromagneticsElectromagnetismMagnetic MaterialsFerrimagnetic MaterialsNeural StimulationNeurostimulationCoil DesignCurrent FocusingElectric Field DistributionExperimental Study DesignRadiofrequency CoilsTranscranial Magnetic Stimulation
A method to improve the focalization of the repetitive transcranial magnetic stimulation figure-eight coil in a magnetic stimulation is presented in this paper. For the purpose of reducing the half width of the distribution curve, while improving the ratio of positive to negative electric field, a shield plate with a window and a magnetic conductor were adopted. The shield plate, which was made of highly conductive copper, focused the magnetic field into a smaller area. The magnetic inductor, which was made of highly permeable soft magnetic ferrite, strengthened the magnetic field. A group of experiments was conducted to validate the focalizing effect. Experimental results showed that the negative peak and the half width of the distribution curve reduced by using the shield plate and the magnetic conductor. Especially for to the Magstim 70 mm double coil, when the shield window was 30 × 60 mm, the ratio of positive to negative electric field could be increased 109%, while the half width of the distribution curve could be reduced about 55%.
11