Validating a New Oscillometric Device for Aortic Pulse Wave Velocity Measurements in Children and Adolescents

90

Citations

32

References

2011

Year

Abstract

Vicorder aPWV values are similar to those obtained by SphygmoCor applanation tonometry. The best agreement between devices was obtained with the path length that most accurately describes the aortic tree. Excellent intra- and interobserver repeatability and ease of measurements make Vicorder appropriate for large multicentre studies in children and adolescents.