Journal of Analytical Atomic Spectrometry · 1998 · 21 citations · 9 references
PharmacyChemistryToxicologyAnalytical ChemistryMass Spectrometric AnalysisSpectroscopic AnalysisGas ChromatographyInstrumental AnalysisIonization TechniquesMass AnalyzersMass SpectrometryCharacteristic MassFragmentation PathwaysInductively Coupled Plasma IonizationMass Spectral FragmentationMass Spectrometric DetectionMass Spectrometric IdentificationPlasma IonizationPlasma-based Ionization
A novel, low power, reduced pressure, helium inductively coupled plasma source (LP/RP-He-ICP), operated at an estimated incident rf power of 12–15 W and pressures ranging from 0.37 to 1.2 mbar, was interfaced with mass spectrometry and coupled with gas chromatography for the separation and detection of three organotin species,viz., tetraethyltin (TET), trimethyl(phenyl)tin (TMPT) and tetrabutyltin (TBT). Limits of detection (as tin) for TET, TMPT and TBT were 0.12, 0.35, and 0.56 pg, respectively. Characteristic mass spectra matching established library spectra were obtained at low nanogram levels for TET,TMPT,TBT and tetramethyltin (TMT). The potential of the source to provide tunable fragmentation was also investigated.
9
David M. Chambers, Jon W. Carnahan, Q. Jin et al. · Spectrochimica Acta Part B Atomic Spectroscopy · 1991 · 31 citations