It is evident that one of the largest sources of imperfectly controlled radiation emanates from dental roentgen-ray machines. This is because they are operated at relatively low kilovoltages and current values for long times of exposure, with no diaphragming of the beam or filtration of it, short focal-film distances and, often, by using standard, relatively slow films. This overdosage can be corrected by employing at least 2 mm Al filtration in the present machines in addition to a lead disphragm and ultra-speed film. To take fall advantage of known physical principles in reducing radiation exposure, higher kilovoltage dental roentgen-ray machines capable of developing 90 or better kv with current capacities of 15 to 20 ana, timers accurate to 1/2th of a second, long-cone techniques, and fastest films are recommended. (auth)