Biomedical Chromatography · 2005 · 60 citations · 11 references
PharmacyBiochemistryAnalytical ChemistryChromatographic AnalysisMass Spectrometric AnalysisBioanalytical Method ValidationDoping Control AnalysisInstrumental AnalysisMass SpectrometryMetabolite IdentificationSteroid Hormone BiochemistryAnabolic Androgenic SteroidsEnzymatic HydrolysisIon Trap Mass SpectrometryLiquid Chromatography-mass SpectrometryLiquid Chromatography-mass Spectrometry AnalysisMass Spectrometric ScreeningSteroid DetectionUrinary Steroid Analysis
A fast and selective LC/MS/MS method for the screening of four anabolic steroids in human urine has been developed and validated. Liquid-liquid extraction with diethyl ether was applied after enzymatic hydrolysis. Analyses were performed on an ion trap mass spectrometer equipped with electrospray ionisation. MS/MS was applied for all compounds. The analytical run time was 11 min. The LOD for all compounds varied between 1 and 10 ng/mL. Left-over A samples, which were declared positive by GC/MS for the presence of 3'-hydroxystanozolol, were assessed using the described method.
11
Antti Leinonen, Tiia Kuuranne, Risto Kostiainen · Journal of Mass Spectrometry · 2002 · 194 citations