A simple, sensitive, and reliable technique has been devised for the radioassay of nonvolatile, water-soluble tritium compounds. The substance to be analyzed is dissolved in an aqueous solution of a thickening agent, preferably sodium <i>O</i>-(carboxymethyl) cellulose or sodium alginate. The solution is placed in a shallow planchet, and after evaporation of the water, the resulting film, which is "infinitely thick" to tritium <i>beta</i> particles, is counted with a 2<i>π</i>, windowless, gas-flow, proportional counter. By means of an empirical factor, determined with a substance of known radioactivity, the counts are converted to microcuries. In a film having a thickness of 0.7 mg/cm<sup>2</sup>, the counting efficiency is about 4 percent; the standard deviation from the mean, obtained in a series of routine measurements, was less than 2 percent. An assay can readily be made with tritium-containing material having 0.01 microcurie of radioactivity. The method, which is applicable to nonvolatile, water-soluble solids, solutions, or liquids, is suitable for routine analyses.

References

4