Measurement of neutron kerma factors for carbon and A-150 plastic at neutron energies of 26.3 MeV and 37.8 MeV

H. Schuhmacher, H.J. Brede, R. Henneck, A. Kunz, J.P. Meulders, P. Pihet, U.J. Schrewe

Physics in Medicine and Biology · 1992 · 14 citations · 27 references

Abstract

Kerma factors for carbon and A-150 tissue equivalent plastic were determined in two nearly monoenergetic neutron beams. Kerma was measured with low pressure proportional counters (PCs) with walls made of graphite and A-150 plastic. The neutron fluence was measured with a proton recoil telescope. The corrections for the kerma contributions from low energy neutrons in the beams were determined by time-of-flight measurements with an NE213 scintillation detector and with the Pcs. For neutron energies of (26.3+/-2.9) MeV and (37.8+/-2.5) MeV the kerma factors for carbon were (35.5+/-3.0) pGy cm2 and (42.6+/-3.9) pGy cm2, respectively, and for A-150 they were (75.6+/-5.4) pGy cm2 and (79.0+/-5.6) pGy cm2. At these energies the kerma factor ratio for ICRU muscle tissue to A-150 calculated on the basis of the new kerma factors, is 0.93.

References

27