Rise in chronic defibrillation thresholds in nonthoracotomy implantable defibrillator.

Ferdinand J. Venditti, David T. Martin, George A. Vassolas, Susan Bowen

Circulation · 1994 · 78 citations · 28 references

DOIFull text

Open access

Abstract

We conclude that there is a significant rise in DFT at 2 and 6 months in this TCD system. Although the chronic threshold remained well within the available energy range of the pulse generator, this observation has important implications for implantation guidelines, programming, and future pulse generator development for TCD patients.

References