Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine · 2013 · 18 citations · 43 references
Composite PlateEngineeringBiomechanical AnalysisMechanical EngineeringBiomedical EngineeringInfrared ThermographyOsteoporosisOrthopaedic SurgerySkeletal TraumaOrthopaedicsStress ShieldingBone HealthStructural Health MonitoringTraditional Metal PlateSpinal FractureThermographyMechanical PropertiesSurgical StabilizationFracture HealingMedicineHip Implant
Traditional high-stiffness metal plates for Vancouver B1 femur shaft fractures below the tip of a hip implant can cause stress shielding, bone resorption, and implant loosening. This is the first study to compare the biomechanics of a traditional metal plate versus a low-stiffness carbon fibre/epoxy composite plate for this injury. A total hip replacement was implanted in two previously validated intact artificial femurs. Femurs were fitted with either a metal or composite plate and had a 5 mm fracture gap created to simulate a Vancouver B1 shaft fracture. Femurs were cyclically loaded using 5 Hz at 7° of adduction with an average axial load of 800 N (range = 400-1200 N). Overall mechanical stiffnesses and femur and plate thermographic stresses were obtained. Femur/metal plate stiffness (698 N/mm) was only 12% higher than femur/composite plate stiffness (625 N/mm). The femur with the composite plate had 22.7% higher combined average stress compared to the femur with the metal plate, having specific differences of 29.5% (anterior view), 33.9% (posterior view), 1.0% (medial view), and 26.4% (lateral view). The composite plate itself had an average 21.1% reduction in stress compared to the metal plate. The composite plate reduced stress shielding, yet provided adequate stiffness.
43
THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS
Journal of Bone and Joint Surgery · 1972 · 1.7K citations