A Phantom Study of the Geometric Accuracy of Computed Tomographic and Magnetic Resonance Imaging Stereotactic Localization with the Leksell Stereotactic System

Cheng Yu, Michael L.J. Apuzzo, Chi‐Shing Zee, Zbigniew Petrovich

Neurosurgery · 2001 · 99 citations · 21 references

Abstract

The accuracy of target localization for most intracranial lesions during stereotactic radiosurgery can be achieved within the size of a voxel, especially by using the Siemens MRI scanner at current specifications and with a new software patch. However, caution is warranted when imaging peripheral lesions, where the distortion is greatest.

References

21