Publication | Closed Access
Dosimetric comparison of treatment plans based on free breathing, maximum, and average intensity projection CTs for lung cancer SBRT
57
Citations
23
References
2012
Year
Dosimetric characteristics of AIP plans are similar to those of FB plans. Slightly better target volume coverage and significantly lower low-dose region (≤30 Gy) in lung was observed in MIP plans. The decrease in low-dose region in lung was mainly caused by the change of lung volume contoured on two datasets rather than the differences of dose distribution between AIP and MIP plans. Compare with AIP datasets, FB datasets were more prone to significant image artifacts and MIP datasets may overestimate or underestimate the target volume when the target is closer to the denser tissue, so AIP seems favorable for planning and dose calculation for lung SBRT.
| Year | Citations | |
|---|---|---|
Page 1
Page 1