JAMA · 1987 · 24 citations · 5 references
PharmacyPharmacoeconomicsAnalytical ChemistryCost-minimization AnalysisHealth Technology AssessmentAnalytical Method ValidationMedical Technology AssessmentPoint-of-care Diagnostic DevicesClinical Chemistry AssaysClinical Laboratory EvaluationGlucose QuantificationLaboratory CostsLaboratory Test EvaluationPrecisionSupply Cost ManagementThroughput
The accuracy and precision of the Kodak DT-60 tabletop chemistry analyzer were evaluated in the clinical laboratory at the Portland (Ore) Veterans Administration Medical Center, and its operational throughout and cost were estimated. All DT-60 tests that were studied exhibited clinically acceptable precision and, except for the glucose method, accuracy. The accuracy of the glucose method was indeterminate with the available data. Throughput under field conditions was found to be less than half of the manufacturer's claim. The estimated supply cost per test could vary from $1.20 to $5.49 per test, depending on the test type and the number of assays expected to be performed daily. The instrument seems to be accurate, precise, and generally reliable when operated by professional medical technologists.
5
Multilayer film elements for clinical analysis: general concepts.
H.G. Curme, R L Columbus, G M Dappen et al. · Clinical Chemistry · 1978 · 100 citations
Natural Sciences, Bioengineering, Biomedical Engineering +16
Laboratory Test Analysis Near the Patient
Richard Belsey · JAMA · 1986 · 43 citations
Natural Sciences, Internal Medicine, Analytical Chemistry +7
Progress and analytic goals in clinical chemistry.
Gilbert Rk · PubMed · 1975 · 29 citations