Magnetic Resonance in Medicine · 2001 · 54 citations · 23 references
MedicineNuclear MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyAirway ImagingMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingPulmonary ImagingPulmonary Nuclear ImagingMagnetic ResonanceRadiofrequency CoilsRodent Lung Imaging
Using in vivo magnetic resonance microscopy, registered (1)H and hyperpolarized (3)He images of the rat lung were obtained with a resolution of 0.098 x 0.098 x 0.469 mm (4.5 x 10(-3) mm(3)). The requisite stability and SNR was achieved through an integration of scan-synchronous ventilation, dual-frequency RF coils, anisotropic projection encoding, and variable RF excitation. The total acquisition time was 21 min for the (3)He images and 64 min for the (1)H image. Airways down to the 6th and 7th orders are clearly visible. Magn Reson Med 45:365-370, 2001.
23
Lung parenchyma: projection reconstruction MR imaging.
Colleen J. Bergin, John M. Pauly, A Macovski · Radiology · 1991 · 402 citations
MR Imaging with Hyperpolarized <sup>3</sup>He Gas
H. Middleton, Robert D. Black, B. Saam et al. · Magnetic Resonance in Medicine · 1995 · 356 citations