Catheterization and Cardiovascular Diagnosis · 1991 · 14 citations · 4 references
MedicineCardiologySurgeryCardiothoracic SurgeryHemodynamic ManagementInterventional CardiologyBalloon AngioplastyCoronary AngioplastyCoronary Artery SurgeryPercutaneous Coronary InterventionBalloon Inflation PressureCoronary Balloon AngioplastyPolyethylene Terephthalate
A small portion of lesions are refractory to coronary angioplasty even when very high pressures are used. This leads to a failed angioplasty and emergent bypass surgery. We successfully attempted the technique of hugging balloons or two polyethelene terephthalate (PET) balloons inflated simultaneously side by side which successfully dilated a lesion which would not dilate using standard techniques. This technique successfully dilated the lesion as the geometry of two balloons inflated side by side is different from one balloon. Two balloons inflated side by side consists of two outer semi-circles and a central trapezoidal square area. This altered geometric configuration may be important in successfully dilating a lesion refractory to standard dilating techniques. Higher pressures can be attained with smaller balloons as the burst pressure is higher in smaller balloons compared to larger balloons.
4
Higher balloon dilatation pressure in coronary angioplasty
Bernhard Meier, Andreas R. Gruentzig, Spencer B. King et al. · American Heart Journal · 1984 · 66 citations
E. Murat Tuzcu, Conrad Simpfendorfer, Kavita Badhwar et al. · The American Journal of Cardiology · 1988 · 28 citations
Medicine, Cardiology, Surgery +11