Comparison of dosimetric standards of USA and France for HDR brachytherapy
21
Citations
14
References
2005
Year
Hdr BrachytherapyMeasurementTreatment VerificationDosimetric StandardsRadiation MedicineCalibrationBrachytherapyBilateral ComparisonRadiation Therapy PlanningRadiation OncologyNuclear MedicineRadiologyHealth SciencesRadiation TherapyDosimetryNational Dosimetric StandardsRadiobiologyCalibration CoefficientsMedicineAnesthesiology
A bilateral comparison of national dosimetric standards for high dose rate brachytherapy has been conducted between the Laboratoire National Henri Becquerel and the University of Wisconsin Accredited Dosimetry Calibration Laboratory. A complete overview of the methods that are currently in use to establish the two national standards is given. The comparison has been carried out using well-type transfer ionization chambers. Three units have been calibrated in both laboratories, and calibration coefficients have been compared. The discrepancies between the two measurements are within 0.3%.