Investigating CT to CBCT image registration for head and neck proton therapy as a tool for daily dose recalculation

Guillaume Landry, Reinoud Nijhuis, G. Dedes, Josefine Handrack, Christian Thieke, Guillaume Janssens, Jonathan Orban de Xivry, Michael Reiner, Florian Kamp, Jan J. Wilkens,

Medical Physics · 2015 · 132 citations · 31 references

DOIFull text

Open access

Abstract

In this work, the authors generated CBCT based stopping power distributions using DIR of the pCT to a CBCT scan. DIR accuracy was below 1.4 mm as evaluated by the SIFT algorithm. Dose distributions calculated on the vCT agreed well to those calculated on the rpCT when using gamma index evaluation as well as DVH statistics based on the same contours. The use of DIR generated contours introduced variability in DVH statistics.