Electromagnetic field hazards involving adjustable shunt valves in hydrocephalus

Thomas Schneider, Uwe Knauff, Jürgen Nitsch, Raimund Firsching

Journal of neurosurgery · 2002 · 46 citations · 7 references

Abstract

Every surgeon who implants these valves and every patient who receives them should know the possible hazards. The valve selection should be adapted to the environment of the patient. Devices with critical levels of electromagnetic flux that are used in the homes of patients should be replaced by ones with lower magnetic fields. The future construction of these valves should be modified in such a way that their adjustment requires a higher magnetic flux density, so that the valves become less sensitive to unwanted effects from environmental magnetic fields.

References

7