Physics in Medicine and Biology · 1963 · 25 citations · 2 references
EngineeringElectrical EngineeringNuclear EngineeringRadiation DetectionRadiation DosimetrySemiconductor DevicesAbsorbed DoseDetector TechnologyGamma-ray DetectionGamma-ray DetectorsRadiation DetectorsSemiconductor DetectorsJunction Leakage CurrentOpen-circuit VoltageSemiconductor Detector DosimetrySemiconductor Gamma-ray DetectorsSemiconductor Particle DetectorsSilicon Detector DosimetrySilicon DetectorsSilicon Semiconductor Detectors
The properties of irradiated silicon PN junctions are discussed in terms of the reverse bias applied to them. Measurements, which are presented, show that at lower exposure rates the bias should be reduced or removed because the leakage current falls with bias more rapidly than does the radiation induced current. When the bias is removed, the current obtainable from a shunted PN junction is proportional to exposure rate. The open circuit voltage is proportional to the logarithm of the exposure rate over a range of exposure rates. Measurements were made on PN junctions constructed for nuclear particle detectors and on solar cells as well as lithium drifted PIN junctions.
2
Pulse Counters for Gamma-dosimetry
A. R. Jones · Health Physics · 1962 · 19 citations