Analysis of methyl tert.-butyl ether and its degradation products by direct aqueous injection onto gas chromatography with mass spectrometry or flame ionization detection systems

Song Hong, Christian M. Duttweiler, Ann T. Lemley

Journal of Chromatography A · 1999 · 46 citations · 12 references

DOIFull text

Open access

Concepts

Abstract

A method was developed to analyze methyl tert.-butyl ether (MTBE) and its degradation products by gas chromatography with mass spectrometry (GC-MS) or flame ionization detection (FID) with direct aqueous injection. The column had dimensions of 30 m x 0.25 mm with film thickness 0.25 microm and a stationary phase of FFAP (nitroterephthalic acid-modified polyethylene glycol). The optimized GC conditions for non-acid components were as follows: carrier gas flow-rate,l mL/min; oven temperature, 35 degrees C for 5.5 min, ramped to 90 degrees C at 25 degrees C/min, then ramped to 200 degrees C at 40 degrees C/min and held at 200 degrees C for 8 min. The conditions for the acid components were: carrier gas flow-rate, 1 mL/min; oven temperature, 110 degrees C for 2 min, ramped to 150 degrees C at 10 degrees C/min, then ramped to 200 degrees C at 40 degrees C/min. The injection port contained a silanized-glass reverse-cup liner filled with Carbofrit. The minimum concentrations for the linear range for the selective ion monitoring mode were 30 to 100 microg/L, depending on the analytes. The minimum detection limit was 1 mg/L for MTBE and tert.-butanol when using FID. More components could be analyzed with the FFAP-type column than with the cyanopropylphenyl-dimethyl polysiloxane-type column.