Scandinavian Journal of Gastroenterology · 1979 · 95 citations · 11 references
MedicinePhysical MedicineHemodynamic ManagementMedical PhysicsSurgical InstrumentationDelivery SystemsFluid ManagementComplianceSyringe Pumps
In a clinical study of oesophageal manometry with fluid-filled catheters, both a non-perfused system and a perfused system with a syringe pump have been compared to a system with a low-compliance perfusion pump, which served as a reference. Significantly lower values of motility amplitudes, motility derivatives, and partly of LES pressures, and a time delay of up to 0.5 sec of the amplitude maximum were obtained with the non-perfused system and the system with a syringe pump in comparison to the low-compliance system. Since the oesophageal function can be erroneously evaluated by use of a non-perfused system or a perfused system with a syringe pump, such systems cannot be recommended for clinical use.
11
Improved Infusion System for Intraluminal Esophageal Manometry
Ronald C. Arndorfer, John J. Stef, Wylie J. Dodds et al. · Gastroenterology · 1977 · 594 citations · Full text
Quantitation of Lower Esophageal Sphincter Competence
Charles S. Winans, Lauran D. Harris · Gastroenterology · 1967 · 348 citations