Archives of Disease in Childhood · 1989 · 28 citations · 6 references
PharmacyPharmaceuticsPharmacologyDosage Form DesignDrug ClearanceDrug DeliveryNebulizer SystemsNebulizer TherapyDrug DurationNebulization EfficiencyNebulizer Aerosol OutputNebulizer Device DesignNebulizer Performance CharacterizationParticle Size Reduction
The output of the drug and the particle size produced by various nebulisers were assessed, and sensible nebulisation times for sodium cromoglycate calculated. Three of the four nebuliser systems tested produced over 80% of total drug output within the first five minutes of nebulisation. We suggest that five minutes is an optimum nebulisation time for sodium cromoglycate using these nebuliser combinations.
6
K. R. May · Bacteriological Reviews · 1966 · 127 citations · Full text
Evaluation of jet nebulisers for use with gentamicin solution.
Stanton Newman, P. G. D. Pellow, Michelle Clay et al. · Thorax · 1985 · 123 citations · Full text
K R May · Bacteriological Reviews · 1966 · 79 citations