A double plasma gas chromatography injector and detector

O. Naji, A. Manz

Lab on a Chip · 2004 · 13 citations · 7 references

Concepts

Abstract

A direct-current, chip-based plasma has been used for gas sample injection in gas chromatography. A second identical plasma chip has been used as the excitation source for an optical emission detector. The first plasma is normally continually sustained during operation, causing continuous ionisation/fragmentation of the sample, whilst the second plasma records the optical emission downstream. For injection, the first plasma is briefly interrupted, introducing a "plug" of unmodified sample into the system. Injection plug sizes of between 5 and 50 [micro sign]l have been reproducibly obtained, although significantly smaller volumes may be possible with the use of smaller cross-section columns, lower flow rates and/or shorter plasma interruption times.

References

7