Polymer thermoelectret dosimetry

G.W. Fabel, H.K. Henisch

physica status solidi (a) · 1971 · 19 citations · 7 references

Concepts

Abstract

Radiation dosimetry based on thermoelectrets of nylon, mylar, polystyrene, and teflon is described. The effect of storage environment, forming temperature, and forming field strength have been examined for these four materials. The electret polarization was monitored by measuring the surface potential of the electrets by means of a contactless voltmeter probe. The electrets (except those of nylon) were found to be very stable under normal conditions. In the presence of X-ray or thermal neutron radiation they undergo a rapid and cumulative discharge. The total dosage received can be inferred from a comparison of the initial and residual polarizations. For 50 kV X-rays, a dosage of 1 mR is easily detectable by such measurements, and higher sensitivities are believed to be attainable in principle.