Concepedia

Publication | Closed Access

On the sensitivity of common gamma‐index evaluation methods to MLC misalignments in Rapidarc quality assurance

153

Citations

33

References

2013

Year

Abstract

A stricter gamma-index (2%/2 mm) is necessary in order to detect positional errors of the MLC. Nevertheless, the quality assurance procedure of Rapidarc treatment plans must include a thorough examination of where dose discrepancies occur, and professional judgment is needed when interpreting the gamma-index analysis, since even a >90% passing rate using the 2%/2 mm gamma-index criterion does not guarantee the absence of clinically significance dose deviation.

References

YearCitations

Page 1