Kirschner wire fixation for scaphoid fractures: an experimental study in synthetic bones

V. Gokce, Hakan Oflaz, Ali Dülgeroğlu, A. Bora, İzge Günal

Journal of Hand Surgery (European Volume) · 2011 · 11 citations · 12 references

Concepts

Abstract

We have studied the biomechanical stability in vitro of three different Kirschner (K) wire configurations in three types of simulated scaphoid waist fractures. The fractures were created with a saw in Sawbones models. There were three fracture patterns: perpendicular to the long axis of the scaphoid model; and 30° and 20° oblique to that. Two 1.2 mm. K-wires were used in each scaphoid. The three configurations were: parallel; 20° oblique; and crossing. The oblique or crossing configurations of K-wires were the most stable depending on the fracture pattern.

References

12