Critical Care Medicine · 2000 · 21 citations · 15 references
The Deltatrac II MBM-200 metabolic monitor appears accurate for low levels of V(O2) and V(CO2) during mechanical ventilation with F(IO2) levels up to 0.390. With increasing F(IO2) to 0.558, the increase in deviation of V(O2) for single measurements can be of clinical relevance for mechanically ventilated infants and children. The increased fluctuation of delivered F(IO2) on higher levels of F(IO2) is likely the cause of the inaccuracy.
15
Measurement of gas exchange in intensive care
Jukka Takala, Olavi Keinänen, Paula Väisänen et al. · Critical Care Medicine · 1989 · 307 citations