Concepedia

Publication | Closed Access

Effect of Graded Facetectomy on Biomechanics of Dynesys Dynamic Stabilization System

101

Citations

26

References

2011

Year

Abstract

Partial facetectomy had a minimal effect on range of motion on the Dynesys-implanted segment. However, in the case of total facetectomy the motion increased by almost 40% in flexion and by 200% in axial rotation. The higher stresses applied to the screws in Dynesys in specific loadings may lead to higher risk of screw failure in Dynesys than in a generic rigid fixation construct.