Differential absorption lidar for volcanic CO_2 sensing tested in an unstable atmosphere

Manuel Queißer, Mike Burton, Luca Fiorani

Optics Express · 2015 · 40 citations · 27 references

DOIFull text

Open access

Concepts

Abstract

Motivated by the need for an extremely durable and portable instrument to quantify volcanic CO(2) we have produced a corresponding differential absorption lidar (DIAL). It was tested on a volcano (Vulcano, Italy), sensing a non-uniform volcanic CO(2) signal under turbulent atmospheric conditions. The measured CO(2) mixing ratio trend agrees qualitatively well but quantitatively poorly with a reference CO(2) measurement. The disagreement is not in line with the precision of the DIAL determined under conditions that largely exclude atmospheric effects. We show evidence that the disagreement is mainly due to atmospheric turbulence. We conclude that excluding noise associated with atmospheric turbulence, as commonly done in precision analysis of DIAL instruments, may largely underestimate the error of measured CO(2) concentrations in turbulent atmospheric conditions. Implications for volcanic CO(2) sensing with DIAL are outlined.

References

27