Analytical Chemistry · 2005 · 16 citations · 7 references
Natural SciencesMaterials ScienceCeramic MaterialsCeramicsPlasma PhysicsDetector SystemsIon-conducting MaterialsMass SpectrometryIon MobilityIon TransportIon Transport MaterialsIon Transport MechanismsIon Transport ModelingIonic MobilitySwarm ModelingTransport Mechanisms
Measurements of the performance of a miniature, portable 12-mm-diameter, 57-mm-length low-temperature cofired ceramic (LTCC) ion mobility spectrometer drift tube were undertaken to verify models of ion transport and determine the physical shape of the ion "swarms" in the LTCC tube. Simplified two-dimensional Gaussian models of ion swarm shape were fit to measured data to extract geometrical shape parameters. Results indicate that tube-transfer function effects that produce asymmetric ion swarms are minimized in the tube reducing temporal dispersion. Data are presented that illustrate the swarm shape as a function of gate time, electric field magnitude, and total charge in the ion swarm. Characterization and understanding of the ion transport mechanisms and effects that limit the resolution and other performance parameters of miniature IMS drift tubes is essential to the development of practical, robust, portable systems for "first responder" and homeland security missions.
7
Measuring the Resolving Power of Ion Mobility Spectrometers
William F. Siems, Ching‐Yi Wu, Edward E. Tarver et al. · Analytical Chemistry · 1994 · 241 citations
66th meeting of the electrochemical society
Corrosion Science · 1984 · 155 citations
Resolution measurement for ion mobility spectrometry
Souji Rokushika, Hiroyuki Hatano, Michael A. Baim et al. · Analytical Chemistry · 1985 · 133 citations