Journal of Applied Physics · 1969 · 29 citations · 1 references
Natural SciencesElectrical EngineeringMaterials ScienceElectronic MaterialsOptoelectronicsSemiconductor DevicesDetector TechnologySemiconductor DetectorsThermoelectric MaterialsCrystal AxesDetectorsGermanium DetectorsPyroelectric DetectorsPyroelectric MaterialsPyroelectric PropertiesSignal-to-noise RatioThermal Detectors
The theory and performance of two types of pyroelectric TGS detectors working at room temperature, one type having electrodes on the front and back surfaces of the crystal plate and the other type with electrodes on the two-side surfaces, have been investigated. They have the same signal-to-noise (S/N) ratio at high frequencies, but the second type detectors, when used with a high-bias resistor, give a higher signal if they have a long edge parallel to the pyroelectric axis. One TGS detector, modified to give a bandwidth of 100 kHz was compared to a gold-doped germanium detector using a CO2 laser as the source. The S/N ratio of the TGS detector is nearly ⅕ of the germanium detector. The type II detector has also been used to study a HCN laser pulse. The possibilities of pyroelectric detectors working at room temperature appears promising in spite of their intrinsic character of thermal detectors.
1
Infra-red radiation detection by the pyroelectric effect
Jason Ludlow, W H Mitchell, E H Putley et al. · Journal of Scientific Instruments · 1967 · 33 citations