Instrumentation factors affecting variance and bias of quantifying tracer uptake with PET/CT

Robert K. Doot, Joshua Scheuermann, Paul E. Christian, Joel S. Karp, Paul E. Kinahan

Medical Physics · 2010 · 82 citations · 31 references

DOIFull text

Open access

Abstract

Short-term scanner variability is low compared to other sources of error. There are tradeoffs in noise and bias depending on acquisition, processing, and analysis methods. The SD of a new PET measure can be estimated from a known SD if the ratios of available coincident counts between the two PET scanner acquisitions are known and both employ the same ROI definition. Results suggest it is feasible to use PET/CTs from different vendors and sites in clinical trials if they are properly cross-calibrated.