Accuracy and Precision of Selected Wound Area Measurement Methods in Diabetic Foot Ulceration

35

Citations

16

References

2013

Year

Abstract

The Elliptical method gives overestimation up to 33%; thus, it should not be used in applications where the actual wound area is an important parameter (like the prediction of wound healing). The TeleDiaFoS system and the SilhouetteMobile device showed the best accuracy of all used methods; however, the precision of the TeleDiaFoS system was revealed to be higher than the precision of the SilhouetteMobile device. The accuracy and the precision of the Visitrak device are significantly reduced for wounds smaller than 2 cm².