Laboratory Furnace for Studies in Controlled Atmospheres; Melting Points of MgO in a N <sub>2</sub> Atmosphere and of Cr <sub>2</sub> O <sub>3</sub> in N <sub>2</sub> and in Air Atmospheres

R. N. McNally, F. I. PETERS, P. H. Ribbe

Journal of the American Ceramic Society · 1961 · 44 citations · 5 references

Concepts

Abstract

An induction furnace design which readily permits the determination of melting points and basic phase studies is described. The melting point of MgO was redetermined to be 2825°C., and the melting point of Cr 2 O 3 is reported as 2330° and 2315°C. in air and in nitrogen atmospheres, respectively.