MedicineOtolaryngologySurgeryHearing AidsHearing AssessmentOtologyHearing Aid FittingHearing Assistive DevicesHearing ImpairmentHearing LossImplant TechnologySurgical ImplantationAverage GainBilateral Sensorineural Hearing LossClinician AcceptanceIncudostapedial Joint AnatomyPatient Satisfaction AnalysisPatient Satisfaction EvaluationSensorineural Hearing Loss Fitting
OBJECTIVE: To assess the efficacy, morbidity, and patient satisfaction of the SOUNDTEC semi-implantable hearing aid. STUDY DESIGN: Retrospective case review. SETTING: Two tertiary referral centers (the Silverstein Institute and the Atkins Institute). PATIENTS: Sixty-four (four bilateral placements) patients with bilateral moderately severe sensorineural hearing loss: 47 patients had previously worn hearing aids. INTERVENTIONS: After separating the incudostapedial joint, a magnet encased in a titanium canister with a ring was introduced onto the stapes neck. Gelfoam or adipose tissue was used to stabilize the magnet. After 3 months, the external processor was fitted. MAIN OUTCOME MEASURES: At 1 month, audiometric testing was performed and functional gain was assessed. Patient acceptance and implant performance were measured by a visual analogue questionnaire. RESULTS: The device produced an average functional gain of 26 dB. Fifty-five percent of patients complained of magnet movement; this was eliminated in 80% of Silverstein Institute patients when the external processor was worn and fat was used to stabilize the magnet. CONCLUSION: The SOUNDTEC direct device is well tolerated in the majority of patients, with a significant increase in functional gain. Magnet instability and noise were the most frequent complaints and improved with processor placement and anchoring the magnet with fat. This electromagnetic semi-implantable hearing aid confers greater functional gain over conventional hearing aids and reduces occlusion effect and feedback.
5
Middle Ear Electromagnetic Semi-Implantable Hearing Device
J. V. D. Hough, Pamela Matthews, Mark W. Wood et al. · Otology & Neurotology · 2002 · 55 citations
Medicine, Otolaryngology, Surgery +17
Early Clinical Results: SOUNDTEC Implantable Hearing Device Phase II Study
J. V. D. Hough, R. Kent Dyer, Pamela Matthews et al. · The Laryngoscope · 2001 · 43 citations
Determining the Importance of Fundamental Hearing Aid Attributes
Hartmut Meister, Isabel Lausberg, Martin Walger et al. · Otology & Neurotology · 2002 · 36 citations