Pipelined Sampled-Delay Focusing in Ultrasound Imaging Systems

Joo Han Kim, Tae Kyong Song, Song Bai Park

Ultrasonic Imaging · 1987 · 57 citations · 12 references

Concepts

Abstract

A sampled-delay focusing technique was recently proposed by the authors which completely eliminates the use of analog L-C delay lines for beam focusing in completely eliminates the use of analog L-C delay lines for beam focusing in ultrasound B-mode imaging systems. With this approach, the product of sampling rate and maximum time delay is required to be less than unity. To remove this constraint, we propose in this paper a first-in-first-out pipelining technique. This allows one to perform beam steering and dynamic focusing simultaneously on a resolution-cell basis and in a completely digital fashion without the use of analog L-C delay lines.

References

12