Journal of Communications Technology and Electronics · 2011 · 12 citations · 1 references
Natural SciencesBiophysicsPhysicsElectromagnetismMagnetic MaterialsMagnetic Resonance BiophysicsCardiac PhysiologyMagnetic CharacterizationMagnetic Resonance ImagingMagnetic SensorsMagnetic SusceptibilityMagnetic DiagnosticsMagnetic Field MeasurementsSquid Magnetometers
A new series of the MAG-SKAN magnetocardiographic (MCG) diagnostic complexes for the measurement and analysis of the biomagnetic signals generated in human heart is developed, produced, and tested. The complexes are based on the magnetometric units that contain superconducting quantum interference devices (SQUIDs) with liquid-helium cooling and 1–36 channels for the detection of the magnetocardiosignals and the original software for the processing and analysis of the MCG data. More than 300 volunteers (a control group of healthy people and a group of patients suffering from cardiovascular diseases) took part in the complex tests. Several qualitative and quantitative MCG parameters that characterize the electric processes in heart and that can be used to identify the patients from the test groups are determined. The diagnostic sensitivity of the parameters ranges from 85 to 97% and the specificity ranges from 80 to 95%.